Features
- Common Features
- Calibration due reminder informs operator to calibrate the meter regularly
- Stability indicator displays the status of current measurement
- Auto hold function freezes stable, final readings for better viewing and recording
- Manual temperature calibration to rectify the deviations in temperature
- Automatic temperature compensation for accurate measured value over the entire range
- Selectable temperature units (°C or °F) for different application requirements
- Expanded memory stores and records up to 500 readings
- Help message as an operational guide, helps users for quick access to the parameters
- Reset function automatically resumes all the setting back to factory default options
- Store data is transferrable into computer via USB communication interface
- pH mode
- 1-5 points push button calibration to automatically identify pH buffer
- Convenient to select pH buffer standard (USA/NIST/DIN) or custom
- Automatic electrode slope display helps operator whether to replace the sensor
- ORP mode
- 1 point offset calibration to adjust the display value to known standard
- Relative & absolute mV measurements for accurate ORP readings
- Ion concentration mode
- 2-5 points calibration, included with 8 concentration points for selection
- Automatic recognition of ion electrodes, no need to specify type of ion
- mV measurements mode for checking the performance of current ion electrode
- Direct reading of concentration measurements to avoid critical conversation and calculations
- Multiple concentration units to be suitable for various applications
- Conductivity / TDS / Salinity / Resistivity Mode
- 1-5 point for calibration with automatically identification of calibration solutions
- Can select cell constant (0.1/1/10) and normalization temperature (20°C or 25°C)
- TDS conversion factor, seawater, practical salinity measurements mode to meet different water quality measurement requirements
- Linear or pure water compensation improves the measurement accuracy
- Dissolved oxygen mode
- 1-2 points calibration using the air-sealed water & zero oxygen solution
- Manual salinity & barometric pressure compensation for accurate readings
Portable Multi-parameter Water Quality Meter LD-LMPWM-A10
Portable Multi-parameter Water Quality MeterLD-LMPWM-A10is adapted with different modes for measuring quality of water on different parameters such as pH, ORP, Ion, conductivity, TDS, salinity, dissolved oxygen content. Due to its portable design, it is suitable for outdoor measurements i.e. on site water testing, remote area water testing & for research purpose. It is reliable and user friendly.
SKU Code: LD-LMPWM-A10
Also Available in below capacities
pH mode :
- Specifications
- Product Applications
| Memory | Stores up to 500 data sets |
| Output | USB communication interface |
| Weight | 300 g |
| pH mode | |
| Accuracy | ± 2.0 % |
| TDS factor | 0.1 - 1.0 (default 0.5) |
| DO accuracy | ± 0.2 mg/L |
| Resistivity | |
| pH accuracy | ± 0.002 pH |
| ORP accuracy | ±0.2 mV |
| Power supply | 3 × 1.5 V “AA batteries or DC 5 V power adapter |
| TDS accuracy | ± 1 % F.S. |
| Cell constant | K = 0.1, 1, 10 or user-defined |
| Ion accuracy | ± 0.5 % F.S. (monovalent), ± 1 % F.S. (divalent) |
| Calibration due | 0 to 31 days |
| Ion measurement | |
| ORP measurement | |
| Dissolved oxygen | |
| TDS measurement | |
| Measurement modes | Seawater or practical salinity |
| Salinity accuracy | ± 1 % F.S |
| Calibration points | 1 - 2 points |
| Compensation modes | Linear or pre water |
| Pressure correction | 60.0 to 112.5 kPa, 450 - 850 mmHg |
| Salinity correction | 0 to 50 g/L |
| Resistivity accuracy | ± 1 % F.S |
| Salinity measurement | |
| Stability conditions | Low or high |
| pH calibration points | 1 - 5 points |
| % Saturation of Oxygen | 0.0 to 200.0 % |
| Conductivity accuracy | ± 0.5 % F.S. |
| Ion calibration points | 2 - 5 points |
| ORP calibration points | 1 point |
| Resistivity resolution | 0.01, 0.1, 1 |
| Dimension (L × W × H) | 185 × 88 × 32 mm |
| Temperature coefficient | 0.0 to 10.0 % / °C |
| Conductivity measurement | |
| Temperature compensation | 0 to 100 °C, 32 to 212 °F, Manual or Automatic |
| pH calibration solutions | USA, NIST, DIN or user defined |
| Ion calibration solutions | 0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000 ppm, mol/L, mg/L |
| Normalization temperature | 20 °C or 25 °C |
| ORP calibration solutions | mV & relative mV |
| Conductivity calibration points | 1 - 5 points |
| Conductivity calibration solutions | 10 µS / cm, 84 µS / cm, 1413 µS / cm, 12.88mS / cm, 111.8mS / cm |
| The portable multi-parameter water quality meter is used for quality testing of water on different parameters in research and development labs, water resources and plants in nearby or remote area. |